DEXTRUS 4136
Report
- Report Number
- 1028232-2011-01790
- Event Type
- Injury
- Date Received
- August 9, 2011
- Date of Event
- June 17, 2011
- Report Date
- August 2, 2011
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVZ
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE, BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.
BOSTON SCIENTIFIC RECEIVED INFO THAT THIS RIGHT ATRIAL (RA) LEAD EXHIBITED SENSING MEASUREMENT OF 2.2 MILLIVOLTS, IMPEDANCE MEASUREMENTS OF 750 OHMS, AND THRESHOLD MEASUREMENTS OF 1.0 AT 0.5. UPON FURTHER ANALYSIS, THE LEAD WAS FOUND TO HAVE DISLODGED. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS REPOSITIONED. NO ADVERSE PT EFFECTS WERE REPORTED. AS NO FURTHER INFO CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED, SHOULD FURTHER INFO BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEXTRUS 4136 | PACER LEAD | NVZ | BIOTRONIK SE & CO. KG | 358754 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |