FDA Adverse Event Injury Summary report: N

TRIAGE CARDIAC PROFILER KIT

MDR report key: 2213004 · Received August 8, 2011

Report

Report Number
2027969-2011-01756
Event Type
Injury
Date Received
August 8, 2011
Date of Event
June 10, 2011
Report Date
August 8, 2011
Manufacturer
ALERE SAN DIEGO, INC
Product Code
NBC
PMA / PMN Number
030286
Removal / Correction Number
NI
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PATIENT SAMPLES WERE NOT RETURNED FOR INVESTIGATION. PRODUCT SUPPORT TESTED PROFILER LOT #W48418 WITH POSITIVE CONTROLS, CALIBRATORS G AND H AND TWO NORMAL WHOLE BLOOD DONORS FOR TROPONIN RECOVERY. NO FALSE NEGATIVE RESULTS OR ERROR CODES WERE OBSERVED. ALL DATA POINTS WITH CALIBRATOR H WERE WITHIN THE MANUFACTURING 2SD RANGE WITH A PERCENT RECOVERY OF (B)(4), WHICH WAS ALSO WITHIN THE MANUFACTURING FINAL RELEASE SPECIFICATIONS. ALL DATA POINTS WITH CALIBRATOR G WERE ABOVE 0.40 NG/ML AND NO FALSE NEGATIVE RESULTS WERE OBSERVED. PRODUCT SUPPORT DID NOT OBSERVE CUSTOMER'S COMPLAINT OF FALSE NEGATIVE TNI DURING RETAINED TESTING WITH EITHER OF THE POSITIVE CONTROL SAMPLES. PRODUCT SUPPORT COULD NOT RULE OUT ANY SAMPLE SPECIFIC OR ENVIRONMENTAL FACTORS THAT MAY HAVE AFFECTED ANALYTE RECOVERY. (B)(4). THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.

Description of Event or Problem · 1

CALLER REPORTED FALSE NEGATIVE TROPONIN (TNI) ON THE TRIAGE CARDIO PROFILER VERSUS A LAB ANALYZER, DADE DIMENSION. A (B)(6) PT ARRIVED AT THE HOSPITAL WITH COMPLAINTS OF PAIN ON THE SIDE. AN EKG WAS ALSO PERFORMED AND THE PT'S FINAL DIAGNOSIS WAS NON-ST-ELEVATION MYOCARDIAL INFARCTION (NSTEMI).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRIAGE CARDIAC PROFILER KIT CARDIAC MARKER PANEL NBC ALERE SAN DIEGO, INC 97100CP W48418B

Patients

Seq Age Sex Outcome Treatment
1 77 YR Other