PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2011-01995
- Event Type
- Injury
- Date Received
- August 8, 2011
- Date of Event
- July 2, 2011
- Report Date
- July 27, 2011
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE CUSTOMER STATED THAT SHE WAS HOSPITALIZED WITH HIGH BLOOD GLUCOSE LEVELS AND LARGE KETONES. THE REPORTED BLOOD GLUCOSE READING WAS 1900 MG/DL. THE CUSTOMER STATED THAT SHE EXPERIENCED DIZZINESS AND HEADACHES PRIOR TO THE EVENT. THE CUSTOMER HAD BEEN EXPERIENCING HIGH BLOOD GLUCOSE LEVELS FOR THE PAST THREE WEEKS. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT THE CUSTOMER DIDN'T HAVE THE TUBING CLAMP FOR THE HIGH PRESSURE TEST. REVIEWED THE ALARM HISTORY, AND FOUND MULTIPLE MOTOR ERROR ALARMS. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC MINIMED | MMT-522LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Hospitalization |