FDA Adverse Event Malfunction Summary report: N

RADIAL JAW¿ 4

MDR report key: 2212941 · Received August 16, 2011

Report

Report Number
3005099803-2011-02734
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
June 6, 2011
Report Date
July 26, 2011
Manufacturer
BOSTON SCIENTIFIC - COSTA RICA
Product Code
FCL
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THAT THE NEEDLE WAS BENT. FUNCTIONALLY, THE JAWS OPENED, BUT FAILED TO CLOSE PROPERLY DUE TO THE BENT NEEDLE. NO ABNORMALITIES WERE NOTED WITH THE DEVICE RIVETING AND CRIMPING WHICH WERE WITHIN SPECIFICATIONS. THE CONDITION OF THE RETURNED INCIDENT DEVICE WAS CONSISTENT WITH THE COMPLAINT THAT THE NEEDLE WAS BENT. HOWEVER, THE EVALUATION FOUND THAT THE JAWS FAILED TO CLOSE PROPERLY DUE TO THIS ISSUE. THE BEND LIKELY OCCURRED DUE TO ANATOMICAL/PROCEDURAL FACTORS WHICH AFFECTED THE DEVICE INTEGRITY AND LIMITED ITS PERFORMANCE. THEREFORE, THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT.

Additional Manufacturer Narrative · 1

THE PATIENT ID, AGE, GENDER AND WEIGHT ARE UNKNOWN. THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RADIAL JAW 4 BIOPSY FORCEPS DEVICE WAS USED DURING A BIOPSY PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE FORCEPS NEEDLE BENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE PRELIMINARY INVESTIGATION RESULTS; JAWS ARE UNABLE TO CLOSE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RADIAL JAW 4 BIOPSY FORCEPS DEVICE WAS USED DURING A BIOPSY PROCEDURE PERFORMED ON (B)(6), 2011. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE FORCEPS NEEDLE BENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE PRELIMINARY INVESTIGATION RESULTS; JAWS ARE UNABLE TO CLOSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RADIAL JAW¿ 4 FORCEPS, BIOPSY, NON-ELECTRIC FCL BOSTON SCIENTIFIC - COSTA RICA M00513331 13918703

Patients

Seq Age Sex Outcome Treatment
1