FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 2212889 · Received July 12, 2011

Report

Report Number
1824206-2011-03680
Event Type
Malfunction
Date Received
July 12, 2011
Date of Event
June 21, 2011
Report Date
June 23, 2011
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD TECHNICIAN FOUND A BOLT SHEARED OFF ON THE BOTTOM OF THE LIFT ARM. REPAIRS HAVE NOT BEEN COMPLETED.

Description of Event or Problem · 1

THE ACCOUNT STATED THE BED WENT INTO REVERSE TREND LIKE MOTION AND THE PATIENT SLID OUT OF THE BED AND ONTO HER FEET. NO INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED BIRTHING BED HDD HILL-ROM, INC. 3600

Patients

Seq Age Sex Outcome Treatment
1 UNK