FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 2212885 · Received July 12, 2011

Report

Report Number
1824206-2011-03688
Event Type
Malfunction
Date Received
July 12, 2011
Date of Event
June 24, 2011
Report Date
June 24, 2011
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE TREND GAS SPRINGS WERE NOT LOCKING, WHICH ALLOWED THE BED TO RETURN TO LEVEL. THE TECHNICIAN REPLACED THE TREND GAS SPRINGS TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BED WILL NOT STAY IN THE TRENDELENBURG POSITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED BIRTHING BED HDD HILL-ROM, INC. 3600

Patients

Seq Age Sex Outcome Treatment
1