HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-10744
- Event Type
- Malfunction
- Date Received
- August 16, 2011
- Date of Event
- July 27, 2011
- Report Date
- July 27, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAD BEEN DISCARDED AND THE LOT NUMBER WAS UNKNOWN, A BATCH REVIEW COULD NOT BE PERFORMED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW-UP REPORT WILL BE FILED.
(B)(4). THIS COMPLAINT FOR THE SYSTEM ERROR 2240 (AIR IN LINE) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE ROOT CAUSE OF THE COMPLAINT WAS NOT DETERMINED. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER (TSC) REGARDING A SYSTEM ERROR (SE) 2240 WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE, FILL. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE MESSAGE TO THE HOME PATIENT (HP) AND INFORMED HIM TO RECYCLE THE MACHINE. THE TSR INFORMED THE HP TO USE MANUAL BAGS. NO REASON FOR THE ERROR WAS FOUND. THERE WAS PATIENT INVOLVEMENT. PRODUCT SURVEILLANCE CONTACTED THE HOME PATIENT (HP) ON (B)(6) 2011 REGARDING THE SYSTEM ERROR 2240 ALARM. THE HOME PATIENT (HP) STATED THAT THE ISSUE WAS RESOLVED, HOWEVER, THE CAUSE OF THE ALARM REMAINED UNKNOWN. THE HP VERIFIED THAT THERE WERE NO LOOSE CONNECTIONS, DISCONNECTIONS OR DEFECTS ON THE SUPPLIES. PER HP, HE DID NOT RECEIVE ANY INJURY OR MEDICAL INTERVENTION AS A RESULT OF THIS INCIDENT. THE HP STATED THAT HE IS DOING FINE AND CONTINUING THERAPY WITHOUT ISSUES. THE HP STATED THAT HE HAD DISCARDED THE SUPPLIES AFTER THERAPY, AND DID NOT KNOW THE LOT NUMBERS. NO ALLEGATIONS WERE MADE AGAINST ANY OF THE HP'S DIALYSIS PRODUCTS. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | HOMECHOICE |