FDA Adverse Event Malfunction Summary report: N

STERLING ES PTA BALLOON DILATATION CATHETER

MDR report key: 2212565 · Received August 16, 2011

Report

Report Number
2134265-2011-03327
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
July 20, 2011
Report Date
July 25, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K080982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY, A BALLOON RUPTURE OCCURRED. THE 99% STENOSED LESION BEING TREATED WAS LOCATED IN THE CALCIFIED AND MODERATELY TORTUOUS POPLITEAL ARTERY. THE 2MM X 20MM X 142CM STERLING BALLOON CATHETER WAS ADVANCED TO THE TARGET LOCATION. ON THE FIRST INFLATION THE BALLOON REACHED 5ATMS AND RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CURRENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERLING ES PTA BALLOON DILATATION CATHETER CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939134202010 13431865

Patients

Seq Age Sex Outcome Treatment
1