FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2212556 · Received August 16, 2011

Report

Report Number
2531779-2011-05915
Event Type
Malfunction
Date Received
August 16, 2011
Report Date
May 24, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. DURING EVALUATION THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION AND CORROSION WAS OBSERVED ON THE FORCE SENSOR ASSEMBLY.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

EVALUATION REVEALED AN OUT OF CALIBRATION FORCE SENSOR. THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 14 YR