FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 2212535 · Received August 16, 2011

Report

Report Number
2250051-2011-00201
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
August 15, 2011
Report Date
August 16, 2011
Manufacturer
HAMILTON BONADUZ AG
Product Code
MMH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT. THE FSE FOUND A BENT COLLET IN POSITION #8 OF THE INSTRUMENT. THE FSE REPLACED THE COLLET THE FSE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE REAGENT AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTER MMH HAMILTON BONADUZ AG 936400

Patients

Seq Age Sex Outcome Treatment
1