FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2212521 · Received August 16, 2011

Report

Report Number
2531779-2011-05913
Event Type
Malfunction
Date Received
August 16, 2011
Report Date
July 21, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PUMP WAS RETURNED TO ANIMAS. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED; HOWEVER, TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON ALL THE KEYPAD BUTTONS. THE BUTTONS DID NOT HAVE NORMAL SPRING BACK OR CLICK, IT WAS CONFIRMED THAT THE KEYS WERE STICKING WHEN PRESSED AND WHEN THE KEYPAD COVER WAS REMOVED, THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE REPORTER CLAIMED ALL KEYPAD BUTTONS WERE HARD TO PRESS, STICKING AND REQUIRED MULTIPLE PRESSES FOR A RESPONSE. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 17 YR