COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-18978
- Event Type
- Malfunction
- Date Received
- August 16, 2011
- Date of Event
- July 19, 2011
- Report Date
- July 28, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE OF 809:01 WAS NOT REPRODUCED DURING PRODUCT EVALUATION; HOWEVER, THE QUALITY ENGINEER HAS DETERMINED THAT THE ROOT CAUSE OF THIS CONDITION WAS A DEFECTIVE PUMP HEAD MODULE. THE PUMP HEAD MODULE WAS REPLACED IN ORDER TO CORRECT THIS CONDITION. ADDITIONAL INFORMATION: THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. A DEVICE HISTORY REVIEW WAS PERFORMED FINDING NO EXCEPTION, NONCONFORMANCE, OR REWORK THAT OCCURRED DURING THE MANUFACTURING OF THE COMPLAINT LOT OR SERIAL NUMBER.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER PERSONNEL, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE OF 809:01. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT/USER INVOLVEMENT, INJURY OR MEDICAL INTERVENTION. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |