FREESTYLE FREEDOM LITE
Report
- Report Number
- 2954323-2011-04208
- Event Type
- Malfunction
- Date Received
- August 15, 2011
- Date of Event
- August 9, 2011
- Report Date
- October 14, 2011
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL FUNCTIONAL TEST RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. READINGS SIMILAR TO THE READINGS THE CUSTOMER REPORTED WERE FOUND IN THE ETER MEMORY WITHIN 10 MINUTES.
PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW UP REPORT WILL BE FILED ONCE ADDITIONAL INFORMATION IS OBTAINED.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF 479 MG/DL AND 152 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN THE VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.