FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2212279 · Received August 15, 2011

Report

Report Number
2531779-2011-05896
Event Type
Injury
Date Received
August 15, 2011
Date of Event
July 14, 2011
Report Date
July 16, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: NO PRODUCT WAS RETURNED; A RETAIN SAMPLE WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE CARTRIDGE PASSES VISUAL INSPECTION, NO DAMAGE OR DEFECTS WERE OBSERVED TO THE LUER CONNECTION, O-RINGS, PLUNGER, OR CARTRIDGE BODY. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS OR ANYWHERE ELSE IN THE CARTRIDGE. A FORCE TEST WAS PERFORMED WITH NO FAILURES.

Additional Manufacturer Narrative · 1

THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE CARTRIDGE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Description of Event or Problem · 1

THE PATIENT CLAIMED HE WAS ADMITTED TO THE HOSPITAL ON (B)(6) 2011 WITH BLOOD GLUCOSE READING IN THE 400S MG/DL AND WAS TREATED FOR DKA AFTER THERE WAS A LEAKAGE ISSUE FOUND AT THE LEUR CONNECTION OF THE ANIMAS CARTRIDGE. THERE WERE NO ISSUES FOUND WITH THE ANIMAS PUMP, INFUSION SET, AND SITE. THE PRODUCT WAS NOT MISUSED. THE ANIMAS CARTRIDGE IS BEING RETURNED FOR INVESTIGATION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY HAD MEDICAL INTERVENTION FOR DKA AFTER THE CARTRIDGE LEAKAGE ISSUE BEGAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM IR1200/1250/2020/OTP CART B201640

Patients

Seq Age Sex Outcome Treatment
1 35 YR Hospitalization| L| R