FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2212235 · Received August 15, 2011

Report

Report Number
2531779-2011-05893
Event Type
Malfunction
Date Received
August 15, 2011
Report Date
July 18, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED CONTAMINATION UNDER KEYPAD BUTTON CONTACTS. THE BUTTONS INTERMITTENTLY ACTIVATED PUMP FUNCTIONS WHEN PRESSED. UNRELATED TO THE COMPLAINT, THE CHARACTERS ON THE DISPLAY SCREEN WERE DIM AND REDDISH. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE PUMP BUTTONS HAVE NOT ACTIVATED DESIRED PUMP FUNCTIONS AS EXPECTED FOR THE PAST FOUR MONTHS. SHE SAYS, SHE CLEANS THE PUMP WITH ALCOHOL. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 14 YR