FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2212235
·
Received August 15, 2011
Report
- Report Number
- 2531779-2011-05893
- Event Type
- Malfunction
- Date Received
- August 15, 2011
- Report Date
- July 18, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED CONTAMINATION UNDER KEYPAD BUTTON CONTACTS. THE BUTTONS INTERMITTENTLY ACTIVATED PUMP FUNCTIONS WHEN PRESSED. UNRELATED TO THE COMPLAINT, THE CHARACTERS ON THE DISPLAY SCREEN WERE DIM AND REDDISH. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT THE PUMP BUTTONS HAVE NOT ACTIVATED DESIRED PUMP FUNCTIONS AS EXPECTED FOR THE PAST FOUR MONTHS. SHE SAYS, SHE CLEANS THE PUMP WITH ALCOHOL. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |