FDA Adverse Event Malfunction Summary report: N

FREESTYLE FREEDOM LITE

MDR report key: 2212229 · Received August 15, 2011

Report

Report Number
2954323-2011-04205
Event Type
Malfunction
Date Received
August 15, 2011
Date of Event
August 8, 2011
Report Date
September 23, 2011
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW UP REPORT WILL BE FILED ONCE ADDITIONAL INFORMATION IS OBTAINED.

Additional Manufacturer Narrative · 1

THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL FUNCTIONAL TEST RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE NOT FOUND IN THE METER MEMORY WITHIN 10 MINUTES.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF 337 MG/DL, 107 MG/DL, 409 MG/DL, 275, MG/DL AND 124 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN THE VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE FREEDOM LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 1108803

Patients

Seq Age Sex Outcome Treatment
1