FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2212166 · Received July 8, 2011

Report

Report Number
1824206-2011-03623
Event Type
Malfunction
Date Received
July 8, 2011
Date of Event
June 23, 2011
Report Date
June 23, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE MANUAL TREND VALVE BUT THE ISSUE REMAINED. THE TECHNICIAN REPLACED THE HYDRAULIC MANIFOLD TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES WHILE PERFORMING PREVENTIVE MAINTENANCE ON THE BED, THE TECHNICIAN FOUND THE TRENDELENBURG FUNCTION WAS NOT WORKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1