FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2212035 · Received July 8, 2011

Report

Report Number
3008642652-2011-00172
Event Type
Malfunction
Date Received
July 8, 2011
Date of Event
May 30, 2011
Report Date
July 8, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEMS (SERVICE CODE 105; CAN COMM TIMEOUT FAULTS) HAVE BEEN INVESTIGATED. UPON EVALUATION THE TRUNK CABLE CONNECTOR WAS DAMAGED, THUS GENERATING CAN COMM TIMEOUT FAULTS AND SERVICE CODE 105. THE CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY DETERMINED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT. NO PROBLEMS WERE FOUND WITH THE PATIENT'S MONITOR.

Description of Event or Problem · 1

A REVIEW OF A (B)(6) MALE PATIENT'S DOWNLOAD REVEALED SEVERAL CAN COMM TIMEOUT FAULTS AND SERVICE CODE 105. THE PATIENT WAS CONTACTED BY ZOLL CUSTOMER SUPPORT AND ISSUED A REPLACEMENT MONITOR AND ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 68 YR