FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2212024
·
Received July 8, 2011
Report
- Report Number
- 3008642652-2011-00158
- Event Type
- Malfunction
- Date Received
- July 8, 2011
- Date of Event
- June 1, 2011
- Report Date
- July 6, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 204) HAS BEEN CONFIRMED. AS RECEIVED, THE TRUNK CABLE WAS PULLED FROM THE DISTRIBUTION NODE (DN), CAUSING COMMUNICATION ERRORS. THE ROOT CAUSE OF THE DAMAGED CABLE CANNOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY A RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE TRUNK CABLE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PT'S DAUGHTER CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A CODE 204. THE PT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |