FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT ¿

MDR report key: 2212009 · Received August 15, 2011

Report

Report Number
2134265-2011-03295
Event Type
Malfunction
Date Received
August 15, 2011
Date of Event
July 21, 2011
Report Date
July 21, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ES
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: A VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE REVEALED PROXIMAL STENT DAMAGE. STRUTS AT THE PROXIMAL END OF THE STENT WERE MISALIGNED. THE TIP AND BALLOON SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. A PRODUCT MANDREL WAS INSERTED THROUGH THE LUMEN WITH NO RESTRICTIONS NOTED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE IS A COMBINATION PRODUCT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, STENT DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE CALCIFIED CORONARY ARTERY. WHEN ADVANCING A 3.5X16MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS), RESISTANCE WAS ENCOUNTERED. THE DEVICE WAS REMOVED AND IT WAS NOTED THAT THE STRUTS WERE DAMAGED. A DIFFERENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, STENT DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE CALCIFIED CORONARY ARTERY. WHEN ADVANCING A 3.5X16MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS), RESISTANCE WAS ENCOUNTERED. THE DEVICE WAS REMOVED AND IT WAS NOTED THAT THE STRUTS WERE DAMAGED. A DIFFERENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROMUS ELEMENT ¿ STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911316350 14390894

Patients

Seq Age Sex Outcome Treatment
1