APTIMA HPV SCREENING ASSAY
Report
- Report Number
- 2024800-2025-00026
- Event Type
- Injury
- Date Received
- May 30, 2025
- Date of Event
- May 7, 2025
- Report Date
- May 30, 2025
- Manufacturer
- HOLOGIC, INC.
- Product Code
- OYB
- UDI-DI
- 15420045500051
- PMA / PMN Number
- P100042
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
HOLOGIC TECHNICAL SUPPORT (TS) REVIEWED AVAILABLE LOGS FOR BOTH INITIAL TEST WORKLIST (B)(4) AND RETEST WORKLIST (B)(4) AND NOTED NO HARDWARE, SOFTWARE OR CONTAMINATION ISSUES. TS INDICATED THAT THE ISSUE WAS MOST LIKELY DUE TO LOW TARGET OR SAMPLE MISHANDLING. HOLOGIC COMPLETED A RISK ASSESSMENT. THERE IS NO PRODUCT IMPACT. HOLOGIC HAS NOT LEARNED OF ANY ADVERSE PATIENT OUTCOMES DUE TO THIS EVENT.
ON (B)(6) 2025, A CUSTOMER REPORTED TO HOLOGIC RECEIVING DISCREPANT RESULTS USING THE APTIMA HPV (HUMAN PAPILLOMAVIRUS) ASSAY MASTER LOT 907848 ON THE PANTHER INSTRUMENT SN# (B)(6). SAMPLE IDS # (B)(6) WERE PART OF A CLUSTER AND INITIALLY RESULTED HPV POSITIVE IN HPV WL #(B)(6). PER THE CUSTOMER¿S LAB SOP, IF THERE IS A CLUSTER OF THREE HPV POSITIVES CONTAINING RESULTS WITH AN S/CO (SIGNAL-TO-CUTOFF) VALUE < 5, THE SAMPLES WITH S/CO<5 WILL BE RETESTED, AND THE RETESTED RESULT WILL BE REPORTED OUT. UPON RETEST, SAMPLE IDS #(B)(6) RESULTED HPV NEGATIVE, AND THIS RESULT WAS REPORTED OUT. SAMPLE IDS # (B)(6) WERE NOT RETESTED AS THESE SAMPLES RESULTED HPV POSITIVE WITH AN S/CO > 5. NO PATIENT TREATMENT INFORMATION WAS PROVIDED BY THE CUSTOMER. HOLOGIC HAS NOT LEARNED OF ANY ADVERSE PATIENT OUTCOMES DUE TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 27741 | APTIMA HPV SCREENING ASSAY | KIT, RNA DETECTION, HUMAN PAPILLOMAVIRUS | OYB | HOLOGIC, INC. | 907848 | 15420045500051 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |