DRILL, 2.75MM, .066 CANNULATION
Report
- Report Number
- 1220246-2025-02319
- Event Type
- Malfunction
- Date Received
- May 30, 2025
- Date of Event
- May 12, 2025
- Report Date
- July 28, 2025
- Manufacturer
- ARTHREX, INC.
- Product Code
- MDM
- UDI-DI
- 00888867040908
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- 501
Narratives
INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.
ADDITIONAL INFORMATION: D9, G3, H3, H6. COMPLAINT ALLEGATION IS CONFIRMED. UPON VISUAL INSPECTION, IT WAS NOTED THAT THE FLUTES OF THE DRILL WERE CHIPPED, ABRASION MARKS WERE VISIBLE AROUND THE FLUTES AND THE THREE TIPS OF THE FLUTES WERE WORN OUT. FUNCTIONAL TESTING WAS NOT PERFORMED DUE TO DAMAGE TO THE DEVICE. THE MOST LIKELY CAUSE OF FAILURE WAS MISALIGNED INSERTION, WHICH DAMAGED THE DEVICE.
IT WAS REPORTED THAT DURING A LATARJET SURGERY WITH CORACOID BONEBLOCK THE DRILLS WERE BLUNT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. ***UPDATE, KHE, 30-MAY2025 THE DEVICE WAS PART OF THE LOAN SET DLS-GBL-IN-10145474.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 667051 | DRILL, 2.75MM, .066 CANNULATION | MANUAL INSTR, GENERAL SURGICAL | MDM | ARTHREX, INC. | DRILL, 2.75MM, .066 CANNULATION | 022428 | 00888867040908 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |