FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2211676 · Received August 15, 2011

Report

Report Number
2531779-2011-05880
Event Type
Malfunction
Date Received
August 15, 2011
Report Date
July 18, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. UNRELATED TO THE COMPLAINT, THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. THE USER GUIDE INSTRUCTS THE PATIENT THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND / OR THE WATERPROOF FEATURE OF THE PUMP.

Description of Event or Problem · 1

ON (B)(6) 2011, THE PATIENT CONTACTED ANIMAS ALLEGING THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE PATIENT CLAIMED THAT THE PUMP WAS UNRESPONSIVE TO BUTTON PRESSES. THERE IS NO EVIDENCE HOWEVER THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY BLOOD GLUCOSE VALUES, SYMPTOMS, OR TREATMENT SUGGESTIVE OF A SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 1250 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 26 YR