UNICEL DXC 800 SYNCHRON SYSTEM
Report
- Report Number
- 2050012-2011-01540
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Date of Event
- March 4, 2009
- Report Date
- March 9, 2009
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JGS
- PMA / PMN Number
- K042291
- Removal / Correction Number
- Z-0863-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- OTHER
Narratives
THE BCI HOTLINE CUSTOMER SERVICE TECHNICIAN DISCUSSED SAMPLE QUALITY AND EFFECTS OF BACTERIAL CONTAMINATION WITH THE CUSTOMER. THE CUSTOMER INDICATED THAT THEY PERFORMED WEEKLY MANUAL ISE MAINTENANCE AND THEY CULTURE THE ISE SYSTEM IN 3 LOCATIONS ONCE A MONTH AND THAT THE CULTURES WERE FOUND TO BE NEGATIVE. THE BCI HOTLINE REP SENT THE CUSTOMER THE BI-WEEKLY MAINTENANCE INSTRUCTIONS. A BCI FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED TO THE SITE AND PERFORMED TROUBLESHOOTING ON THE SYSTEM. THE FSE REPLACED THE CARBON BRIDGE AND WHILE THIS RESOLVED THE PROBLEM, A CLEAR ROOT CAUSE FOR THE EVENT WAS NOT DETERMINED. THIS IS 1 OF 160 MEDWATCH REPORTS ASSOCIATED WITH THIS EVENT. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS CONDUCTED FROM 01/01/2008 THROUGH 10/23/2010 FOR ADDITIONAL REPORTABLE EVENTS.
THE CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BCI) THAT ERRONEOUSLY LOW SODIUM (NA) AND CHLORINE (CL) RESULTS WERE OBTAINED OUT OF RANGE LOW ON THEIR UNICEL DXC 800 SYNCHRON SYSTEM INSTRUMENT. HOWEVER, THE POTASSIUM (K) RESULTS WERE WITHIN THE ACCEPTABLE RANGE. PRIOR TO THE EVENT, THE ISE (ION-SELECTIVE ELECTRODE) SYSTEM WAS CALIBRATED SUCCESSFULLY AND QC (QUALITY CONTROL) RECOVERED WITHIN THE LAB'S ESTABLISHED RANGES. A PHYSICIAN QUESTIONED THE LOW NA RESULTS AND INDICATED THAT 10 PTS WOULD HAVE TO BE TREATED. THE LAB RERAN THE QC CHECK AGAIN AND FOUND THAT THE NA AND CL QC RECOVERED LOW OUTSIDE OF THE ESTABLISHED RANGES. ALL SAMPLES RUN BETWEEN 2 AM AND 10 AM ON THE DAY OF THE EVENT WHEN REPEATED ON THE LAB'S SECOND DXC ANALYZER AND THE RESULTS FOR NA AND CL WERE CORRECTED BUT THE K RESULTS WERE NOT CORRECTED. THE 160 CORRECTED REPORTS WERE ISSUED OUTSIDE OF THE LAB. THE CUSTOMER PROVIDED 22 EXAMPLES OF PT RESULTS. WHILE THE CUSTOMER DID NOT RECEIVE ANY REPORT OF ANY ADVERSE EVENT OR PT INJURY ASSOCIATED WITH THIS EVENT, IT IS KNOWN THAT THERE WAS CHANGE TO PT TREATMENT IN AT LEAST 10 CASES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 800 SYNCHRON SYSTEM | JGS | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |