FDA Adverse Event
Malfunction
Summary report: N
HS III PROXIMAL SEAL SYSTEM 3.8MM
MDR report key: 2211518
·
Received August 5, 2011
Report
- Report Number
- 2242352-2011-01155
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Date of Event
- August 3, 2011
- Report Date
- August 3, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K080169
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR INVESTIGATION. WE WILL CONTINUE PURSUING THE DEVICE BEING RETURNED BY THE CUSTOMER. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED LOT NUMBER. THERE WERE NO NON CONFORMANCE ISSUE(S) WITH THIS PRODUCTION LOT. INTERNAL FILE NUMBER- (B)(4).
Description of Event or Problem · 1
THE HOSPITAL REPORTED THAT DURING PREPARATION FOR A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL WAS NOTICED TO BE CRACKED DURING LOADING. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS. MAQUET CARDIOVASCULAR ANTICIPATES RETURN OF THE PRODUCT IN QUESTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HS III PROXIMAL SEAL SYSTEM 3.8MM | CLAMPLESS BEATING HEART | DXC | MAQUET CARDIOVASCULAR, LLC | HSK-3038 | 25037379 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |