FDA Adverse Event Injury Summary report: N

OTHER

MDR report key: 2211474 · Received August 15, 2011

Report

Report Number
2531779-2011-05875
Event Type
Injury
Date Received
August 15, 2011
Report Date
September 9, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

BRAND NAME: IR1200/IR1250/2020/OTP BATTERY CAP. COMMON DEVICE NAME: BATTERY CAP. MODEL #: IR1200/IR1250/2020/OTP. (B)(6). CUSTOMER SUPPORT REVIEWED THE PUMP WITH THE REPORTER AND DETERMINED THAT THE BATTERY CAP WAS DAMAGED. THE ALLEGED MALFUNCTION IS NOT LIKELY TO CAUSE AN ADVERSE EVENT. THE ISSUE IS GENERALLY OBVIOUS AND DETECTABLE BY THE USER. THEREFORE, THE DETECTABLE FLAW MAY PREVENT THE USER FROM CONTINUING USE OF THE DEVICE OR CAUSE A CONDITION THAT MAKES CONTINUED USE OF THE DEVICE IMPOSSIBLE. THE OWNER'S BOOKLET INSTRUCTS THE USER TO CALL ANIMAS CUSTOMER SERVICE IF THE USER SUSPECTS THAT THE PRODUCT IS DAMAGED. THE FAMILY MEMBER WAS ADVISED TO TAKE THE PATIENT OFF THE PUMP AND USE A BACK UP PLAN. THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, EVALUATION WILL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT (B)(6) 2010, THE PATIENT WAS TAKEN TO THE EMERGENCY ROOM FOR ELEVATED BLOOD GLUCOSE (BG). THERE WAS NO SPECIFIC BG VALUE GIVEN. AT THE TIME OF THE CALL TO ANIMAS CUSTOMER SUPPORT (CS), THE FAMILY MEMBER STATED THAT THE PATIENT'S BG WAS BETWEEN 350 AND 400 MG/DL. THERE WERE NO REPORTED SIGNS OR SYMPTOMS OF HYPERGLYCEMIA. THE FAMILY MEMBER STATED THAT THE BATTERY CAP THREADS WERE STRIPPED AND THE PUMP HAD BEEN REBOOTING. CUSTOMER SUPPORT ADVISED THE FAMILY MEMBER THAT THE PATIENT SHOULD BE DISCONNECTED FROM THE PUMP AND PLACED ON A BACK-UP PLAN UNTIL THE BATTERY CAP COULD BE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OTHER INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 12 YR Required Intervention