TI MATRIX LOCKING CAP
Report
- Report Number
- 2530088-2011-00508
- Event Type
- Injury
- Date Received
- August 9, 2011
- Date of Event
- July 7, 2011
- Report Date
- July 11, 2011
- Manufacturer
- SYNTHES BRANDYWINE
- Product Code
- NKB
- PMA / PMN Number
- K100952
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION COULD NOT BE COMPLETED, NO CONCLUSION COULD BE DRAWN AS NO DEVICE WAS RETURNED AND NO LOT NUMBER WAS PROVIDED.
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
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TI MATRIX LOCKING CAP | MATRIX LOCKING CAP | NKB | SYNTHES BRANDYWINE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Required Intervention | LOCKING CAPS| RODS| SCREWS |