OTHER
Report
- Report Number
- 2531779-2011-05874
- Event Type
- Injury
- Date Received
- August 15, 2011
- Report Date
- September 9, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
BRAND NAME: IR1200/IR1250/2020/OTP BATTERY CAP. COMMON DEVICE NAME: BATTERY CAP. MODEL #: IR1200/IR1250/2020/OTP. (B)(6). CUSTOMER SUPPORT REVIEWED THE PUMP WITH THE REPORTER AND DETERMINED THAT THE BATTERY CAP WAS DAMAGED. THE ALLEGED MALFUNCTION IS NOT LIKELY TO CAUSE AN ADVERSE EVENT. THE ISSUE IS GENERALLY OBVIOUS AND DETECTABLE BY THE USER. THEREFORE THE DETECTABLE FLAW MAY PREVENT THE USER FROM CONTINUING USE OF THE DEVICE OR CAUSE A CONDITION THAT MAKES CONTINUED USE OF THE DEVICE IMPOSSIBLE. THE OWNER'S BOOKLET INSTRUCTS THE USER TO CALL ANIMAS CUSTOMER SERVICE IF THE USER SUSPECTS THAT THE PRODUCT IS DAMAGED. THE FAMILY MEMBER WAS ADVISED TO TAKE THE PATIENT OFF THE PUMP AND USE A BACK UP PLAN. THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, EVALUATION WILL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
A FAMILY MEMBER REPORTED THAT IN (B)(6) 2010 THE PATIENT EXPERIENCED BLOOD GLUCOSE (BG) AROUND 500 MG/DL AND WAS TAKEN TO THE HOSPITAL. SHE STATED THAT THERE WAS NO REASON FOUND FOR THE ELEVATED BG, BUT SHE BELIEVED IT MAY HAVE BEEN DUE TO THE BATTERY CAP NOT SECURING TO THE PUMP BECAUSE OF STRIPPED THREADS. THE FAMILY MEMBER DENIED ANY SITE/SET ISSUES. CUSTOMER SUPPORT ADVISED THE FAMILY MEMBER TO PUT THE PATIENT ON A BACK-UP PLAN UNTIL THE BATTERY CAP COULD BE REPLACED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THIS COMPLAINT IS BEING REPORTED DUE TO THE PATIENT'S ALLEGED HYPERGLYCEMIC EVENT WHILE USING INSULIN PUMP THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OTHER | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR |