FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX PTCA DILATATION CATHETER

MDR report key: 2211419 · Received August 15, 2011

Report

Report Number
2134265-2011-03302
Event Type
Malfunction
Date Received
August 15, 2011
Date of Event
July 13, 2011
Report Date
July 21, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURED. THE 85% STENOSED DE NOVA LESION WAS LOCATED IN THE SEVERELY CALCIFIED PROXIMAL RIGHT CORONARY ARTERY. A 3.50X15MM NC QUANTUM APEX BALLOON CATHETER WAS SELECTED FOR PRE-DILATION AND ADVANCED TO THE LESION WITHOUT RESISTANCE. ON THE FIRST INFLATION, THE BALLOON RUPTURED AT 6 ATMS AFTER BEING INFLATED FOR ABOUT 5 SECONDS. THE DEVICE WAS REMOVED INTACT .THE PHYSICIAN THEN ADVANCED A 3.50X10MM FLEXTOME OVER THE WIRE (OTW) CUTTING BALLOON. ON THE 1ST INFLATION, THE CUTTING BALLOON RUPTURED AT 4ATMS AFTER BEING INFLATED FOR ABOUT 5 SECONDS. THE PHYSICIAN ATTEMPTED TO ADVANCE A SECOND 4.00X15MM FLEXTOME OTW CUTTING BALLOON; HOWEVER, ON THE FIRST INFLATION THE CUTTING BALLOON RUPTURED AT 6ATMS AFTER BEING INFLATED FOR ABOUT 5 SECONDS. BOTH OF THE CUTTING BALLOONS WERE REMOVED INTACT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX PTCA DILATATION CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912415350 14392848

Patients

Seq Age Sex Outcome Treatment
1 89 YR