ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-05871
- Event Type
- Malfunction
- Date Received
- August 15, 2011
- Date of Event
- September 2, 2010
- Report Date
- September 2, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/27/2010 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE WORN BUT INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. A LEAK TEST WAS PERFORMED AND A KEYPAD LEAK WAS OBSERVED.
THE PATIENT REPORTED THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS BECAME INTERMITTENTLY UNRESPONSIVE APPROXIMATELY TWO MONTHS PRIOR TO REPORTING THE ISSUE TO CUSTOMER SUPPORT. SHE STATED THAT SHE WORE THE PUMP IN A SWIMMING POOL ON ONE OCCASION. SHE DENIED DAMAGE TO THE KEYPAD, AND STATED THAT THE KEYPAD BUTTONS HAD "LIMITED CLICK AND SPRING-BACK."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |