FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2211320 · Received August 15, 2011

Report

Report Number
2531779-2011-05871
Event Type
Malfunction
Date Received
August 15, 2011
Date of Event
September 2, 2010
Report Date
September 2, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/27/2010 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE WORN BUT INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. A LEAK TEST WAS PERFORMED AND A KEYPAD LEAK WAS OBSERVED.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS BECAME INTERMITTENTLY UNRESPONSIVE APPROXIMATELY TWO MONTHS PRIOR TO REPORTING THE ISSUE TO CUSTOMER SUPPORT. SHE STATED THAT SHE WORE THE PUMP IN A SWIMMING POOL ON ONE OCCASION. SHE DENIED DAMAGE TO THE KEYPAD, AND STATED THAT THE KEYPAD BUTTONS HAD "LIMITED CLICK AND SPRING-BACK."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 62 YR