FDA Adverse Event Malfunction Summary report: N

ENDOWRIST SP

MDR report key: 22111976 · Received May 30, 2025

Report

Report Number
2955842-2025-23055
Event Type
Malfunction
Date Received
May 30, 2025
Date of Event
May 9, 2025
Report Date
May 9, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874114292
PMA / PMN Number
K182371
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE MEDIUM-LARGE CLIP APPLIER INSTRUMENT WAS COULD NOT VERIFY THE CUSTOMER REPORTED EVENT. THE INSTRUMENT WAS TESTED ON AN IN-HOUSE SYSTEM. THE INSTRUMENT PASSED THE RECOGNITION AND ENGAGEMENT TESTS. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIP TIPS OPENED AND CLOSED PROPERLY. THE INSTRUMENT PASSED THE CLIP TEST. THE INSTRUMENT WAS FULLY FUNCTIONAL. THE COMPLAINT WAS NOT CONFIRMED BY FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE 6 MM SINGLE PORT MEDIUM-LARGE CLIP APPLIER INSTRUMENT TO PERFORM FAILURE ANALYSIS. THE INSTRUMENT WAS TESTED ON AN IN-HOUSE SYSTEM. THE INSTRUMENT PASSED THE RECOGNITION AND ENGAGEMENT TESTS. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIP TIPS OPENED AND CLOSED PROPERLY. THE INSTRUMENT PASSED THE CLIP TEST. THE INSTRUMENT WAS FULLY FUNCTIONAL.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 6 MM SINGLE PORT MEDIUM-LARGE CLIP APPLIER INSTRUMENT WAS NOT CLIPPING. NO FRAGMENT FELL INTO THE PATIENT. THE USER COMPLETED THE PROCEDURE AND NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
642822 ENDOWRIST SP MEDIUM-LARGE CLIP APPLIER NAY INTUITIVE SURGICAL, INC 430005-57 K10230817 0005 00886874114292

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES