FDA Adverse Event Malfunction Summary report: N

7700

MDR report key: 2211156 · Received August 8, 2011

Report

Report Number
9680959-2011-01687
Event Type
Malfunction
Date Received
August 8, 2011
Date of Event
June 28, 2011
Report Date
August 8, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE CABLES TO THE CONNECTOR INSIDE THE C-ARM WERE REPLACED. THE SYSTEM WAS FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A WHITE SCREEN ON THE LEFT MONITOR, AND THAT THE X-RAY LIGHT STAYED ON AFTER RELEASING THE FOOT SWITCH. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7700 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7700

Patients

Seq Age Sex Outcome Treatment
1