FDA Adverse Event Malfunction Summary report: N

SYMPHONY

MDR report key: 2211135 · Received August 8, 2011

Report

Report Number
9610579-2011-00083
Event Type
Malfunction
Date Received
August 8, 2011
Date of Event
July 18, 2011
Report Date
July 19, 2011
Manufacturer
SORIN CRM
Product Code
NVZ
PMA / PMN Number
P950029
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE EVENT CONCERNS A DEVICE THAT WAS MANUFACTURED AND USED OUTSIDE THE UNITED STATES. ANALYSIS IS PENDING.

Description of Event or Problem · 1

REPORTEDLY ON (B)(6) 2011, THE PHYSICIAN OBSERVED THAT THE ESTIMATED LONGEVITY WAS 130 MONTHS. HOWEVER ON (B)(6) 2011 THE ESTIMATED LONGEVITY WAS 157 MONTHS. THE PHYSICIAN REQUESTED AN EXPLANATION. PRELIMINARY ANALYSIS HAS SHOWED THAT THE PHYSICIAN HAS CHANGED THE PACING PARAMETERS ON (B)(6) 2011 AND THAT BOTH ESTIMATIONS WERE IN ACCORDANCE WITH THE PROGRAMMED PARAMETERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYMPHONY NVZ SORIN CRM SYMPHONY DR 2550 S050906

Patients

Seq Age Sex Outcome Treatment
1