FDA Adverse Event Malfunction Summary report: N

ACTIVA PC

MDR report key: 2211122 · Received August 8, 2011

Report

Report Number
3004209178-2011-06156
Event Type
Malfunction
Date Received
August 8, 2011
Date of Event
June 1, 2011
Report Date
October 12, 2011
Manufacturer
MDT PUERTO RICO OPERATIONS CO
Product Code
MHY
PMA / PMN Number
P960009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED FROM THE HCP REPORTED THAT THE CAUSE OF THE EVENT WAS A CHANGE IN THE PATIENT'S MEDICATION (ADDED COMTAN TO SINEMET). IT WAS UNLIKELY THAT ANY PART OF THE DEEP BRAIN STIMULATION SYSTEM WAS AN ISSUE. THE ELECTRODE (#2) WITH THE HIGH IMPEDANCES WAS NOT BEING USED IN THE THERAPY, BUT DID PRODUCE SIDE EFFECTS (ARM PAIN, DIFFICULTY SPEAKING, AND FACIAL PULLING) WHEN TESTED. AS OF THE LAST VISIT, THE PATIENT WAS DOING WELL WITH THE STIMULATION PROGRAM AND MEDICATION. THE PATIENT DID REQUIRE AND ER VISIT AND A CT SCAN OF THE HEAD FOLLOWING THE FALL, BUT DID NOT REQUIRE AN EXTENDED STAY ADMISSION. THE PATIENT RECOVERED WITHOUT SEQUELAE. THE HCP NOTED THAT THE LEFT STIMULATION PROGRAM WAS CONTROLLING THE LEFT SIDE OF THE BODY, AND THE RIGHT STIMULATION PROGRAM WAS CONTROLLING THE RIGHT SIDE OF THE BODY, EVEN THOUGH THE LEAD CONFIGURATION WAS APPROPRIATELY SET UP.

Description of Event or Problem · 1

IT WAS REPORTED THAT ALL COMBINATIONS WITH #2 ARE >40000 OHMS. A LOSS OF THERAPEUTIC EFFECT WAS REPORTED. PATIENT REPORTED THEY MAY STILL BE GETTING SOME STIMULATION BENEFIT. THE PATIENT FELL TWO WEEKS AGO AND HIT HIS HEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACTIVA PC STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR MHY MDT PUERTO RICO OPERATIONS CO 37601 NA

Patients

Seq Age Sex Outcome Treatment
1 74 YR EXPLANTED:| EXTENSION: MODEL 7482, LOT# NHU153590V| IMPLANTED:| EXPLANTED:| LEAD: MODEL 3387S, LOT# V107750| EXTENSION: MODEL 7482, LOT# NHU153589V| PROGRAMMER: MODEL 37642, LOT# NJZ106931N| ADAPTER: MODEL 64002, LOT# N248399| EXPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| LEAD: MODEL 3387S, LOT# V107750| IMPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED: