FDA Adverse Event
Malfunction
Summary report: N
6800
MDR report key: 2211096
·
Received August 8, 2011
Report
- Report Number
- 1720753-2011-20785
- Event Type
- Malfunction
- Date Received
- August 8, 2011
- Date of Event
- July 25, 2011
- Report Date
- August 8, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE SERVICE REPRESENTATIVE RESEATED THE BOARD, CHECKED THE POWER SUPPLY FOR PROPER VOLTAGE, FLASHED THE NODES AND RELOADED THE CALIBRATION FILES. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT PERFORM FLUOROSCOPY. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 6800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 6800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |