ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-05864
- Event Type
- Malfunction
- Date Received
- August 14, 2011
- Report Date
- July 15, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP (B)(4) HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(4).
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: PRIOR INVESTIGATION FOUND THERE WAS NO EVIDENCE OF DAMAGE TO THE BATTERY COMPARTMENT OR CAP AND NO CORROSION WAS FOUND IN THE BATTERY COMPARTMENT. ALL PUMP CURRENT DRAWS WERE FOUND TO BE WITHIN SPECIFICATIONS. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT ALARMS. THE PUMP WAS OPENED FOR INVESTIGATION AND NO FURTHER MOISTURE OR CORROSION DAMAGE WAS FOUND. THE POWER CIRCUIT WAS INSPECTED AND NO DEFECTS WERE FOUND.
IT WAS ALLEGED THAT THE PUMP EMITTED FREQUENT REPLACE BATTERY ALARMS (EVERY TWO WEEKS, INTERMITTENTLY) AND THAT MULTIPLE BATTERIES FELT HOT WHEN REMOVED. A FAMILY MEMBER STATED THAT THE PATIENT USED 12 BATTERIES (B)(6) 2011. SHE REPORTED THAT THE PATIENT USED ULTIMATE LITHIUM L91, THE BATTERY TYPE WAS CORRECTLY PROGRAMMED, THERE WERE NO UNCONFIRMED ALARMS, AND THE BATTERY CAP FIT SECURELY TO THE PUMP. THE FAMILY MEMBER CONFIRMED THAT NO ONE WAS INJURED RELATED TO THE HOT BATTERIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR |