BECKMAN COULTER ¿ COULTER ® LH750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2011-01177
- Event Type
- Malfunction
- Date Received
- August 15, 2011
- Date of Event
- July 11, 2011
- Report Date
- July 13, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
SAMPLE WAS DRAWN INTO BD EDTA TUBE. THE INSTRUMENT IS CURRENTLY WITHIN QC SPECIFICATIONS WITH RESPECT TO CONTROLS AND REPRODUCIBILITY. FSE VERIFIED THE INSTRUMENT AND FOUND NO PROBLEMS. THE SAMPLES RUN BEFORE AND AFTER THE SAMPLE IN QUESTION LOOKED GOOD WITH NO IRREGULARITIES. RAW DATA ANALYSES: FOR THE HIGH EO RUN, ALL POPULATIONS SHIFTED TO THE RIGHT SIDE ON THE DF1 HISTOGRAM. SOME OF THE NEUTROPHILS WERE CLASSIFIED AS EOSINOPHILS BECAUSE OF THIS SHIFTING. DATA DISCONTINUITY WAS DETECTED DURING THE ASPIRATION, BUT THE ALGORITHM DID NOT SET A SYSTEM OR SUSPECT FLAG. ALL THREE RUNS WITH ELEVATED PLT RESULTS SHOWED LOW-END INTERFERENCE PATTERNS AND WERE FLAGGED R. ROOT CAUSE FOR THE ERRONEOUS DIFF WAS ALL POPULATIONS SHIFTED TO THE RIGHT SIDE ON THE DF1 HISTOGRAM AND SOME NEUTROPHILS WERE CLASSIFIED AS EOSINOPHILS. ALL THREE RUNS WITH ELEVATED PLT RESULTS SHOWED LOW-END INTERFERENCE PATTERNS AND WERE FLAGGED R. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) IN REGARDS TO AN ERRONEOUS HIGH EO AND LOW NE RESULTS WITHOUT ANY DIFF FLAGS ON ONE RUN (OUT OF FIVE) OF A PATIENT SPECIMEN WHEN COMPARED TO MANUAL DIFFERENTIAL. THREE ADDITIONAL RUNS OF THE SAME SPECIMEN ALSO GAVE ERRONEOUS HIGH PLT WITH INSTRUMENT GENERATED FLAGS AS COMPARED TO PLT ESTIMATE FROM SMEAR AND PATIENT HISTORY. ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LAB. A MANUAL SMEAR WAS DONE BY A TECH AND CONFIRMED THE PLT RESULT OF 56. THE MANUAL DIFFERENTIAL RESULTS SHOWED LOWER EO%/HIGHER NE%. MANUAL DIFFERENTIAL AND PLT ESTIMATE WAS REPORTED AS CORRECT. NO DEATH, INJURY OR CHANGE TO PATIENT TREATMENT IS ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BECKMAN COULTER ¿ COULTER ® LH750 HEMATOLOGY ANALYZER | AUTOMATED DIFFERENTIAL CELL COUNTER | GKZ | BECKMAN COULTER INC. | LH750 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |