UNICEL® DXC SYNCHRON ® CLINICAL SYSTEMS
Report
- Report Number
- 2050012-2011-04272
- Event Type
- Malfunction
- Date Received
- August 14, 2011
- Date of Event
- July 12, 2011
- Report Date
- July 14, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JGS
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
THE SAMPLE IS SERUM. NO ADDITIONAL SAMPLE INFORMATION WAS PROVIDED. QC WAS PROVIDED FROM (B)(6) 2011 ONLY AND LOW LEVEL NA QC WAS OUT OF RANGE LOW. QC WAS REPEATED 7 HOURS LATER AND IT WAS WITHIN LABORATORY ESTABLISHED RANGE. PER CUSTOMER, CO2 ALK BUFFER LINE 30 AND DAMPER LOOKED CLOUD. A BEC FIELD SERVICE ENGINEER (FSE) REPLACED THE CHLORIDE TIP AND CARBON DIOXIDE ELECTRODES. FSE VERIFIED PERFORMANCE.
THE CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) IN REGARDS TO ERRONEOUS LOW SODIUM (NA) RESULT ON 1 PATIENT SAMPLE GENERATED BY THE UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEMS. THE ERRONEOUS RESULT WAS NOT REPORTED OUT OF THE LABORATORY. THE LABORATORY WAS ALERTED OF A LOW ANION GAP AND THE SAMPLE WAS REPEATED ON AN ALTERNATE DXC AND HIGHER RESULTS WAS OBTAINED AND REPORTED. PATIENT TREATMENT WAS NOT IMPACTED BY THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC SYNCHRON ® CLINICAL SYSTEMS | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE | JGS | BECKMAN COULTER INC. | DXC 600PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |