FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2211012 · Received August 13, 2011

Report

Report Number
2531779-2011-05863
Event Type
Malfunction
Date Received
August 13, 2011
Report Date
July 17, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(6). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE WORN BUT INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT THE KEYPAD BUTTONS HAVE BEEN INTERMITTENTLY UNRESPONSIVE FOR THE PAST TWO TO THREE MONTHS. SHE CONFIRMED THAT THE KEYPAD IS INTACT AND THAT THE PUMP WAS NOT EXPOSED TO MOISTURE. THE FAMILY MEMBER NOTED THAT THE KEYPAD BUTTONS STICK AT TIMES. SHE STATED THAT THE PATIENT WEARS THE PUMP ON HER BELT AND DOES NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 11 YR