UNICEL® DXC SYNCHRON ® CLINICAL SYSTEMS
Report
- Report Number
- 2050012-2011-04271
- Event Type
- Malfunction
- Date Received
- August 13, 2011
- Date of Event
- July 13, 2011
- Report Date
- July 13, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
THE SAMPLES WERE SERUM. NO ADDITIONAL SAMPLE INFORMATION WAS PROVIDED. PER CUSTOMER, QC RUN PRIOR TO THE EVENT WAS WITHIN LABORATORY ESTABLISHED RANGES. QC WAS NOT RUN AFTER THE EVENT. A BEC FIELD SERVICE ENGINEER REPLACED A CRACKED ELECTROLYTE INJECTION CUP VALVE AND VERIFIED PERFORMANCE.
A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) IN REGARDS TO ERRONEOUS CHLORIDE RESULTS GENERATED ON THE UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEMS. THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF LABORATORY; HENCE PATIENT TREATMENT WAS NOT IMPACTED. WHEN ANION GAPS FLAGGED THE SAMPLES, THE SAMPLES WERE RERUN ON AN ALTERNATE INSTRUMENT, AND THOSE RESULTS REPORTED. THE CUSTOMER COULD NOT PROVIDE ACTUAL RESULTS OR VERBAL EXAMPLES. THE NUMBER OF SAMPLES IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC SYNCHRON ® CLINICAL SYSTEMS | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE | JJE | BECKMAN COULTER INC. | DXC 800PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |