UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-04062
- Event Type
- Malfunction
- Date Received
- August 12, 2011
- Date of Event
- July 13, 2011
- Report Date
- July 13, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
BEC CTS (CUSTOMER TECHNICAL SUPPORT) ADVISED CUSTOMER TO HAVE PROPER PROTECTIVE EQUIPMENT (PPE) ON AND USE INSTRUMENT COMMAND TO SHUTDOWN SYSTEM. A SERVICE REQUEST WAS GENERATED AND FIELD SERVICE ENGINEER (FSE) WENT ON-SITE (B)(6) 2011 AND FOUND MODULAR CHEMISTRY (MC) SAMPLE PROBE PLUGGED. FSE REPLACED THE PROBE AND MC TRANSDUCER AND THE ISSUE WAS RESOLVED. CALIBRATION WAS PERFORMED AND REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES AND RESULTS MET PUBLISHED PERFORMANCE SPECIFICATIONS. THE BEC INTERNAL DENTIFER FOR THIS EVENT IS (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING A MC (MODULAR CHEMISTRY) OBSTRUCTION DETECTION TRANSDUCER FAILURE AND UNRECOVERABLE MC OBSTRUCTION IN THE UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM. PER CUSTOMER, THERE WERE ALSO LIQUID ON TOP OF THE INSTRUMENT MODULAR COVERS AND THE SAMPLE CAROUSEL COVER. THE CUSTOMER WAS UNABLE TO DETERMINE SOURCE OF THE LEAK OR THE TYPE OF LIQUID BUT STATED IT WAS CLEAR LIQUID, POSSIBLY WATER OR WASH SOLUTION. NO INJURY OR EXPOSURE WAS REPORTED AND NO ERRONEOUS RESULTS WERE GENERATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |