FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2210791 · Received August 12, 2011

Report

Report Number
1423500-2011-10681
Event Type
Malfunction
Date Received
August 12, 2011
Date of Event
July 29, 2011
Report Date
July 29, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE DEVICE FAILED THE HOMECHOICE RITE ELECTRICAL TEST AND FAILED THE RITE FUNCTIONAL TEST. RITE ELECTRICAL TEST FAILURE WAS CONFIRMED BY REVIEW OF THE RITE ELECTRICAL TEST REPORT. THE CAUSE OF THE RITE ELECTRICAL TEST FAILURE WAS DETERMINED TO BE CAUSED BY FLUID INGRESS INTO THE COMPRESSOR AND MANIFOLD. THE HC DEVICE TO BE FORWARDED TO SERVICE AND THE DEVICE TO BE SCRAPPED. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, THE PRODUCT ANALYSIS LABORATORY DETERMINED THE HC MACHINE SYSTEM FAILED RETURNED INSTRUMENT TEST/EVALUATION TESTING DUE TO A FAILURE OF THE EARTH LEAKAGE CURRENT TEST. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1