FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2210710 · Received August 12, 2011

Report

Report Number
2531779-2011-05847
Event Type
Malfunction
Date Received
August 12, 2011
Report Date
July 15, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED. THE PUMP WAS INTERMITTENTLY UNRESPONSIVE TO ALL KEYPAD BUTTON PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS FOUND UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

ON (B)(6) 2011, THE PATIENT CONTACTED ANIMAS ALLEGING THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE PATIENT CLAIMED THAT THE PUMP WAS UNRESPONSIVE TO DOWN BUTTON PRESSES. THERE IS NO EVIDENCE HOWEVER THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY BLOOD GLUCOSE VALUES, SYMPTOMS, OR TREATMENT SUGGESTIVE OF A SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 11 YR