FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2210636 · Received August 12, 2011

Report

Report Number
2531779-2011-05844
Event Type
Malfunction
Date Received
August 12, 2011
Report Date
July 17, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PUMP WAS RETURNED TO ANIMAS. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED; HOWEVER, TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON ALL KEYPAD BUTTONS. IN ADDITION, THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION FOUND UNDER ALL KEY CONTACTS. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT CLAIMED THE OK KEYPAD BUTTON REQUIRED MULTIPLE PRESSES FOR A RESPONSE. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 46 YR