FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM

MDR report key: 2210582 · Received August 10, 2011

Report

Report Number
2953200-2011-01402
Event Type
Injury
Date Received
August 10, 2011
Date of Event
July 26, 2011
Report Date
July 26, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL, RESULTS: GRAFT MIGRATION, ENDOLEAK. RESULTS/CONCLUSIONS: DISEASE PROGRESSION AND AORTIC NECK DILATATION.

Description of Event or Problem · 1

AN ANEURX ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 46 MONTHS AGO. ANEURYSM AND VESSEL MORPHOLOGY FROM THE TIME OF IMPLANT ARE UNK. IT WAS REPORTED THAT THE PT WAS BEING EVALUATED FOR KIDNEY STONES, WHEN IT WAS NOTED THAT THE ANEURX BIFURCATED STENT GRAFT HAD MIGRATED DISTALLY AND WAS FOUND TO BE 30 MM BELOW THE RENAL ARTERIES WITH A PROXIMAL TYPE I ENDOLEAK EVIDENT. THE AORTIC NECK HAD 30 TO 40 DEGREE ANGULATION WITH A DIAMETER OF 27 MM PROXIMALLY AND 28 MM DISTALLY. THE STENT GRAFT MIGRATION WAS ATTRIBUTED TO DISEASE PROGRESSION AND AORTIC NECK DILATATION. THE PHYSICIAN ELECTED TO IMPLANT TWO 32 MM ENDURANT CUFFS, WHICH SUCCESSFULLY RESOLVED TO THE ENDOLEAK AND MIGRATION. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR NA V00023073

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention