TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO
Report
- Report Number
- 2953200-2011-01421
- Event Type
- Injury
- Date Received
- August 10, 2011
- Date of Event
- July 12, 2011
- Report Date
- July 12, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVAL, RESULTS: GRAFT MIGRATION, ENDOLEAK, CONVERSION TO OPEN REPAIR. RESULTS/CONCLUSIONS: DISEASE PROGRESSION AND NO AORTIC NECK PRESENT.
A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 18 MONTHS AGO. ANEURYSM AND VESSEL MORPHOLOGY FROM THE TIME OF IMPLANT ARE UNK. IT WAS REPORTED THAT THE PT PRESENTED APPROXIMATELY 1 MONTH AGO WITH A PROXIMAL TYPE I ENDOLEAK, AND A CT DEMONSTRATED THAT THE STENT GRAFT WAS FOUND TO HAVE MIGRATED DISTALLY ALL THE WAY INTO THE ANEURYSM SAC. IT WAS NOTED THAT THE ANEURYSM HAS GROWN SIZE. THERE IS ILIAC FIXATION BILATERALLY ALL THE WAY TO THE HYPOGASTRIC ARTERIES. THE PT'S LAST FOLLOW-UP WAS 5 MONTHS AGO, AND NO MIGRATION WAS NOTED AT THE TIME. THE GRAFT MIGRATION WAS ATTRIBUTED TO DISEASE PROGRESSION, WHICH HAS RESULTED IN AN ABSENCE OF AN AORTIC NECK. THE PHYSICIAN ELECTED TO CONVERT THE PT TO OPEN REPAIR. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00414647 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Required Intervention |