FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO

MDR report key: 2210580 · Received August 10, 2011

Report

Report Number
2953200-2011-01421
Event Type
Injury
Date Received
August 10, 2011
Date of Event
July 12, 2011
Report Date
July 12, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL, RESULTS: GRAFT MIGRATION, ENDOLEAK, CONVERSION TO OPEN REPAIR. RESULTS/CONCLUSIONS: DISEASE PROGRESSION AND NO AORTIC NECK PRESENT.

Description of Event or Problem · 1

A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 18 MONTHS AGO. ANEURYSM AND VESSEL MORPHOLOGY FROM THE TIME OF IMPLANT ARE UNK. IT WAS REPORTED THAT THE PT PRESENTED APPROXIMATELY 1 MONTH AGO WITH A PROXIMAL TYPE I ENDOLEAK, AND A CT DEMONSTRATED THAT THE STENT GRAFT WAS FOUND TO HAVE MIGRATED DISTALLY ALL THE WAY INTO THE ANEURYSM SAC. IT WAS NOTED THAT THE ANEURYSM HAS GROWN SIZE. THERE IS ILIAC FIXATION BILATERALLY ALL THE WAY TO THE HYPOGASTRIC ARTERIES. THE PT'S LAST FOLLOW-UP WAS 5 MONTHS AGO, AND NO MIGRATION WAS NOTED AT THE TIME. THE GRAFT MIGRATION WAS ATTRIBUTED TO DISEASE PROGRESSION, WHICH HAS RESULTED IN AN ABSENCE OF AN AORTIC NECK. THE PHYSICIAN ELECTED TO CONVERT THE PT TO OPEN REPAIR. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00414647

Patients

Seq Age Sex Outcome Treatment
1 68 YR Required Intervention