VASOVIEW HEMOPRO 2
Report
- Report Number
- 2242352-2011-01171
- Event Type
- Malfunction
- Date Received
- August 8, 2011
- Date of Event
- July 27, 2011
- Report Date
- August 1, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K101274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR INVESTIGATION. WE WILL CONTINUE PURSUING THE DEVICE BEING RETURNED BY THE CUSTOMER. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WERE NO NON CONFORMANCE ISSUE (S) WITH THIS PRODUCT LOT. INTERNAL FILE NUMBER - (B)(4).
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE SURGEON REPORTED THAT THE VASOVIEW HEMOPRO 2 "CLAMP SHOULD BE SENT". THE COMPLAINT FORM STATED "COMPLAINT FROM OUR CUSTOMER, THE CLAMP SHOULD BE BENT ON THE TOP SIDE. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS. MAQUET CARDIOVASCULAR ANTICIPATES RETURN OF THE PRODUCT IN QUESTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO 2 | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC | VH-4000 | 25038455 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |