EXTENSION, 30 CM
Report
- Report Number
- 1627487-2011-06049
- Event Type
- Injury
- Date Received
- August 10, 2011
- Date of Event
- July 12, 2011
- Report Date
- July 12, 2011
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- K991784
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVAL: RESULTS: VISUAL EXAMINATION OF THE LEAD EXTENSION REVEALED THAT ALL OF THE WIRES FROM THE HEADER WERE PULLED BACK AND EXTRACTED TO THE STRAIN RELIEF. NO FUNCTIONAL TESTING WAS PERFORMED DUE TO THE BROKEN WIRES. MOST LIKELY THE STRESS WAS INDUCED WHILE THE DEVICE WAS IMPLANTED. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MFR REPORT # 1627487-2011-06048. THE PT'S (B)(6) SCS SYSTEM INCLUDED AN IPG, LEAD AND LEAD EXTENSION. IT WAS REPORTED THAT THE PT WAS WITHOUT STIMULATION. BEFORE THE STIMULATION CEASED, THE PT'S THERAPY WAS ALLEGEDLY INTERMITTENT, AND THE IPG HAD BECOME DIFFICULT TO RECHARGE. SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2011 TO ADDRESS THIS MATTER. PRIOR TO THE PROCEDURE, THE PT'S IPG REGISTERED A LOW BATTERY WARNING. ATTEMPTS TO RECHARGE THE DEVICE WERE REPORTEDLY UNSUCCESSFUL. INTRAOPERATIVE TESTING OF THE PT'S LEAD REVEALED NORMAL IMPEDANCE READINGS WHEN TESTED INDEPENDENTLY. HOWEVER, WHEN THE LEAD WAS TESTED IN CONJUNCTION WITH THE LEAD EXTENSION, HIGH IMPEDANCE MEASUREMENTS WERE OBSERVED. FURTHER INTERROGATION REVEALED THAT THE PT'S EXTENSION WAS BROKEN. AS SUCH, BOTH THE PT'S IPG AND LEAD EXTENSION WERE REPLACED. NO FURTHER ISSUES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EXTENSION, 30 CM | SPINAL CORD STIMULATION LEAD EXTENSION | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3383 | 84794 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | SCS LEAD: MODEL UNK| IMPLANTED: |