FDA Adverse Event Injury Summary report: N

EXTENSION, 30 CM

MDR report key: 2210497 · Received August 10, 2011

Report

Report Number
1627487-2011-06049
Event Type
Injury
Date Received
August 10, 2011
Date of Event
July 12, 2011
Report Date
July 12, 2011
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
GZB
PMA / PMN Number
K991784
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVAL: RESULTS: VISUAL EXAMINATION OF THE LEAD EXTENSION REVEALED THAT ALL OF THE WIRES FROM THE HEADER WERE PULLED BACK AND EXTRACTED TO THE STRAIN RELIEF. NO FUNCTIONAL TESTING WAS PERFORMED DUE TO THE BROKEN WIRES. MOST LIKELY THE STRESS WAS INDUCED WHILE THE DEVICE WAS IMPLANTED. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT # 1627487-2011-06048. THE PT'S (B)(6) SCS SYSTEM INCLUDED AN IPG, LEAD AND LEAD EXTENSION. IT WAS REPORTED THAT THE PT WAS WITHOUT STIMULATION. BEFORE THE STIMULATION CEASED, THE PT'S THERAPY WAS ALLEGEDLY INTERMITTENT, AND THE IPG HAD BECOME DIFFICULT TO RECHARGE. SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2011 TO ADDRESS THIS MATTER. PRIOR TO THE PROCEDURE, THE PT'S IPG REGISTERED A LOW BATTERY WARNING. ATTEMPTS TO RECHARGE THE DEVICE WERE REPORTEDLY UNSUCCESSFUL. INTRAOPERATIVE TESTING OF THE PT'S LEAD REVEALED NORMAL IMPEDANCE READINGS WHEN TESTED INDEPENDENTLY. HOWEVER, WHEN THE LEAD WAS TESTED IN CONJUNCTION WITH THE LEAD EXTENSION, HIGH IMPEDANCE MEASUREMENTS WERE OBSERVED. FURTHER INTERROGATION REVEALED THAT THE PT'S EXTENSION WAS BROKEN. AS SUCH, BOTH THE PT'S IPG AND LEAD EXTENSION WERE REPLACED. NO FURTHER ISSUES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXTENSION, 30 CM SPINAL CORD STIMULATION LEAD EXTENSION GZB ST. JUDE MEDICAL - NEUROMODULATION 3383 84794

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SCS LEAD: MODEL UNK| IMPLANTED: