FDA Adverse Event Malfunction Summary report: N

3002 SECURE II

MDR report key: 2210223 · Received July 11, 2011

Report

Report Number
1831750-2011-06882
Event Type
Malfunction
Date Received
July 11, 2011
Date of Event
June 13, 2011
Report Date
June 13, 2011
Manufacturer
STRYKER MEDICAL
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULT: POWER CORD PLUG MISSING GROUND PRONG. THIS USER FACILITY SERVES AS THE HEALTH CARE PROVIDER FOR ONE OF THE STATE PRISONS. AS A RESULT, IT HAS BEEN REPORTED THAT PTS INTENTIONALLY DAMAGE VARIOUS DEVICE COMPONENTS WITH UNRESTRAINED APPENDAGES. ADDITIONALLY PTS ARE REGULARLY RESTRAINED IN MANNERS THAT CONFLICT WITH THE DEVICE'S INSTRUCTIONS FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED BY SERVICE REPORT THAT THE POWER CORD PLUG WAS MISSING THE GROUND PRONG. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 3002 SECURE II A/C POWERED HOSPITAL BED FNL STRYKER MEDICAL 3002 NA

Patients

Seq Age Sex Outcome Treatment
1