FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 22102027 · Received May 28, 2025

Report

Report Number
9617229-2025-08783
Event Type
Injury
Date Received
May 28, 2025
Date of Event
November 7, 2024
Report Date
May 28, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
5060191607278
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS OF PAIN, CAPSULAR CONTRACTURE AND ANXIETY-PRODUCT/PROCEDURE WAS RECEIVED ON (B)(6) 2025, WITH LOT NUMBER 2251210. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINT CODES ARE: PAIN: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. CAPSULAR CONTRACTURE: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. ANXIETY-PRODUCT/PROCEDURE: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. AS PER THE INVESTIGATION PROCEDURE, DEFORMATION WAS OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED. THE EVENT OF "CAPSULAR CONTRACTURE, BAKER GRADE IV" IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE IV

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED TEXTURED TO SMOOTH EXCHANGE. PATIENT REPORTED PAIN AND DISCOMFORT. LATER HEALTHCARE PROFESSIONAL REPORTED BILATERAL GRADE 4 CAPSULAR CONTRACTURE. THIS RECORD IS FOR THE LEFT SIDE. DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
587958 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2251210 5060191607278

Patients

Seq Age Sex Outcome Treatment
1 41 YR Female Required Intervention